Control of Components and Reagents, continued

CFR Title 21 Section 606.65 regulates how often reagents that are used for processing blood and compatibility testing must be quality control tested. This section also requires that reagents be used according to the manufacturer's instructions.
The US Code of Federal Regulations sets minimum standards for reagent potency. To ensure that this potency remains intact, a representative sample of each lot of the reagents or solutions in Table 2 must be tested on a regularly scheduled basis by methods described in the facility's Standard Operating Procedures manual to determine its capacity to perform as required.
Table 2. Reagent/Solution Testing Frequency.10
Reagent or Solution
Frequency of Testing
Anti-human globulin
Each day of use
Blood grouping reagents
Each day of use
Lectins
Each day of use
Antibody screening and reverse grouping cells
Each day of use
Enzymes
Each day of use
Hepatitis test reagents
Each run
Syphilis serology reagents
Each run
Reagent storage
Reagents must be used within their dating period. Supplies and reagents that do not bear an expiration date should be stored so that the oldest is used first and should be used in a manner consistent with instructions provided by the manufacturer.
Most blood grouping reagents and cells are labeled to be stored between 2–8°C. When blood components and reagents are stored in the same blood refrigerator, the low alarm needs to be set to activate at 2.5°C rather than 1.5°C.
10. Code of Federal Regulations. Part 606-Current Good Manufacturing Practice for Blood and Blood Components, Subpart D, Section 606.65 "Supplies and reagents." Food and Drug Administration, Updated 09 Apr 2024. Available at:https://www.ecfr.gov/current/title-21/chapter-I/subchapter-F/part-606/subpart-D/section-606.65.